G3516B Engines Caterpillar


Oxygen Sensor - Calibrate

Usage:

G3516B 7EZ
System Operation Description:

Periodic calibration of the oxygen sensor is required in order to ensure accurate readings and optimum engine performance. Refer to the Operation And Maintenance Manual for the maintenance interval schedule. The engine must be stopped and the sensor must be removed. The sensor must be placed in uncontaminated air in order to calibrate the sensor. The air must be free of exhaust.

During the calibration, the Electronic Control Module (ECM) first establishes the reference for zero percent of oxygen by reading the sensor after the sensor has cooled for two minutes. The ECM energizes the heater in the sensor for five minutes. Last, the ECM reads the percent of oxygen in the surrounding air. The surrounding air is assumed to be clean with approximately 20.95 percent oxygen. The calibration offset and the calibration range are determined from these two points.

Test Step 1. Check the Atmospheric Conditions

  1. Determine the ambient temperature, relative humidity, and barometric pressure.

  2. Use Table 1 to determine whether your relative humidity is within the permissible range for calibration of the oxygen sensor.

    1. Find the row for your ambient temperature.

      If the temperature is less than 12 °C (54 °F), use the 12 °C (54 °F) row.

      If the temperature is more than 38 °C (100 °F), do not calibrate the sensor.

    2. Find the column for your barometric pressure.

    3. Find the relative humidity in the cell that intersects your column and your row.

      If your relative humidity is equal to the humidity or greater than the humidity that is listed in the Table, do not calibrate the sensor.

      If the humidity is 90 percent or more, do not calibrate the sensor.

Table 1
Ambient Temperature, Barometric Pressure, and Relative Humidity for Calibration of the Oxygen Sensor 
Ambient Temperature  Barometric Pressure 
96 kPa (13.9 psi) 97 kPa (14.7 psi)  99 kPa (14.4 psi)  101 kPa (14.6 psi)  102 kPa (14.8 psi)  104 kPa (15 psi)  105 kPa (15.2 psi)  107 kPa (15.5 psi)  109 kPa (15.8 psi)  110 kPa (16 psi)  112 kPa (16.2 psi)  113 kPa (16.4 psi) 
Relative Humidity
12 °C (54 °F)  90  90  90  90  90  90  90  90  90  90  90  90 
14 °C (57 °F)  85  85  85  85  90  90  90  90  90  90  90  90 
16 °C (61 °F)  75  75  75  75  80  80  80  80  85  85  85  85 
18 °C (64 °F)  65  65  65  70  70  70  70  75  75  75  75  75 
20 °C (68 °F)  60  60  60  60  60  65  65  65  65  65  65  70 
22 °C (72 °F)  50  50  55  55  55  55  55  60  60  60  60  60 
24 °C (75 °F)  45  45  45  50  50  50  50  50  50  55  55  55 
26 °C (79 °F)  40  40  40  45  45  45  45  45  45  45  50  50 
28 °C (82 °F)  35  35  40  40  40  40  40  40  40  40  45  45 
30 °C (86 °F)  35  35  35  35  35  35  35  35  35  40  40  40 
32 °C (90 °F)  30  30  30  30  30  30  35  35  35  35  35  35 
34 °C (93 °F)  25  25  30  30  30  30  30  30  30  30  30  30 
36 °C (97 °F)  25  25  25  25  25  25  25  25  25  30  30  30 
38 °C (100 °F)  20  25  25  25  25  25  25  25  25  25  25  25 

Expected Result:

The temperature is less than 38 °C (100 °F). The ambient relative humidity is within the permissible range for calibration of the oxygen sensor.

Results:

  • OK - The ambient conditions are acceptable. Proceed to Test Step 2.

  • Not OK - The humidity is not acceptable.

    Repair: Do not calibrate the sensor until the humidity is acceptable.

    STOP

Test Step 2. Check for Diagnostic Codes

  1. Switch the 16 amp circuit breaker ON. Set the engine control to the STOP mode.

  2. Access the "Logged Diagnostic Codes" screen on the Caterpillar Electronic Technician ET. If the following codes are logged, clear the codes:

    • 1086-09 Oxygen Sensor Element not communicating on link

    • 1086-13 Oxygen Sensor Element calibration required

    • 1087-03 Oxygen Buffer Module short to +batt

    • 1087-08 Oxygen Buffer Module noisy signal

    • 1088-05 Oxygen Sensor Power Supply open circuit

    • 1088-06 Oxygen Sensor Power Supply short to ground

  3. Access the "Diagnostic Override" screen of Cat ET. Select "Diagnostics/Diagnostic Tests/Override Parameters".

  4. Turn ON the "Oxygen Sensor Override".

  5. Allow a minimum of 90 seconds for any diagnostic codes to be generated.

  6. Access the "Logged Diagnostic Codes" screen again. Look for the following diagnostic codes again: 1086-09, 1086-13, 1087-03, 1087-08, 1088-05 and 1088-06.

Expected Result:

None of the above diagnostic codes are logged.

Results:

  • No Codes - The circuit for the oxygen sensor is OK. Proceed to test Step 3.

  • Logged 1086-09, 1087-03 or 1087-08 - A 1086-09 1087-03 or 1087-08 diagnostic code is active.

    Repair: Refer to Troubleshooting, "Oxygen Sensor Signal".

    STOP

  • Logged 1088-05 or 1088-06 Diagnostic Code - There is a problem with the supply voltage for the oxygen buffer.

    Repair: Refer to Troubleshooting, "Oxygen Sensor Buffer Supply".

    STOP

Test Step 3. Calibrate the Oxygen Sensor

  1. Set the engine control to the OFF/RESET mode. Switch the 16 amp circuit breaker OFF.

    ------ WARNING! ------

    During operation, the oxygen sensor may reach temperatures that exceed 700 °C (1292 °F). Severe personal injury and property damage will result from contact with a hot oxygen sensor.

    Do not touch the sensor during engine operation, calibration, or testing. Allow the sensor to cool before moving the sensor. Wear gloves that resist heat. Do not place the sensor on or near any flammable material or any surface that can be damaged by high temperatures.


  2. Disconnect the oxygen sensor from the oxygen buffer.

  3. Allow the sensor to cool sufficiently for removal. Then remove the sensor from the exhaust system.

  4. Connect the oxygen sensor to the oxygen buffer. Suspend the sensor in uncontaminated air.

  5. Switch the 16 amp circuit breaker ON. Set the engine control to the STOP mode.

  6. Access the "Service/Calibrations/Oxygen Sensor" screen of Cat ET. Use Cat ET to start the calibration.

    Approximately 12 minutes are required for the calibration.

Expected Result:

Successful calibration is indicated by Cat ET.

Results:

  • OK - The calibration was successful. Proceed to Test Step 4.

  • Not OK - The calibration was not successful. Cat ET will display a diagnostic code.

    Repair: Troubleshoot the diagnostic code. Check the connectors for the oxygen sensor and the oxygen buffer. Repeat the calibration procedure. If the procedure is not successful after three attempts, obtain a new oxygen sensor and repeat the calibration procedure.

    STOP

Test Step 4. Verify the Calibration of the Sensor

  1. Access the "Diagnostic Override" screen of Cat ET. Select "Diagnostics/Diagnostic Tests/Override Parameters".

  2. Turn ON the "Oxygen Sensor Override" parameter.

Expected Result:

The "Actual Oxygen" reading stabilizes to approximately 20.95 percent in about 30 seconds.

Results:

  • OK - The calibration of the oxygen sensor is verified. Allow the sensor to cool before you install the sensor.

    STOP

  • Not OK - The calibration of the oxygen sensor is not verified.

    Repair: Repeat the calibration procedure.

    STOP

Caterpillar Information System:

JF Series Starting Motor Starting Motor - Assemble
Electronic Modular Control Panel II+ (EMCP II+) GSC CID 0447 - FMI 12 (for EUI Engines)
2000/07/01 Correction to The Pocket Guide, SEBD0402 Is Available as a Reference for Routing and Installing Wire Harness Assemblies {1408}
Alternator - Delco 21SI Insulated Screw - Check
EMCPII+ for Gas Engines GSC CID 0175 - FMI 02
C-10, C-12, 3406E, C-15, C-16 and C-18 Engines - Electronic PTO Installation and Application PTO Configurations - Test
3500B and 3500B High Displacement Generator Set Engines Configuration Parameters
3306C Truck Engine Increased Engine Oil Temperature - Inspect
C7 Engines for Caterpillar Built Machines Excessive Bearing Wear - Inspect
3306C Truck Engine Excessive Engine Oil Consumption - Inspect
3500B Engines for Caterpillar Built Power Modules Shutdown Notification Output
3306C Truck Engine Water Temperature Regulator - Test
Alternator - Delco 21SI Rectifier - Test
2000/10/23 Correction to The Pocket Guide, SEBD0402, Is Available as a Reference for Routing and Installing Wire Harness Assemblies {1408}
EMCPII+ for Gas Engines GSC CID 0334 - FMI 03
2000/07/01 Identify OEM Selling Dealer On-line {1000}
EMCPII+ for Gas Engines GSC CID 0175 - FMI 03
3126 Truck Engine (Military) Welding on Engines with Electronic Controls
3608 Engine Air Starting Motor Silencer - Remove and Install
3306C Truck Engine Slave Piston Lash - Adjust
EMCPII+ for Gas Engines GSC CID 0859 - FMI 03
Electronic Modular Control Panel II+ (EMCP II+) GSC CID 0447 - FMI 12 (for MUI Engines)
3126B and 3126E On-highway Engines Powertrain Data Link Circuit - Test
EMCPII+ for Gas Engines GSC CID 0175 - FMI 04
Back to top
The names Caterpillar, John Deere, JD, JCB, Hyundai or any other original equipment manufacturers are registered trademarks of the respective original equipment manufacturers. All names, descriptions, numbers and symbols are used for reference purposes only.
CH-Part.com is in no way associated with any of the manufacturers we have listed. All manufacturer's names and descriptions are for reference only.