Using Caterpillar Electronic Technician (Cat® ET)
The Caterpillar Electronic Technician (Cat® ET) service tool is used to access the calibration procedures. For more information, refer to the manual that was provided with Cat® ET.
Similar procedures for calibration may also be performed through the Monitoring System. See Systems Operation, RENR8068, "Calibration Mode".
Machine calibration is used to compensate for the differences in the mechanical and electrical components. A calibration should be performed if the electronic control module (ECM) has been replaced. Calibration should also be performed if related components have been replaced or removed. Perform all calibrations that are applicable.
Note: Be sure that the hydraulic oil temperature is at least 50 °C (122 °F) before performing calibrations.
Accessing Calibrations
Connect Cat® ET to the machine at the Service Connector. See Illustration 1. After communication has been established, Cat® ET will list the ECMs that are on the machine.
Illustration 1 | g03477479 |
Cat® ET Service Connector Location (Behind Operator Seat) (1) Service Connector (2) Rear of Operator Seat (shifted forward for access) (3) Main Backup Switch (4) Backup Engine Speed Selector Switch |
Note: The ECM must detect specific setup conditions before beginning the calibration procedure. If a setup step is not met, the calibration procedure is aborted.
- Select the Machine ECM.
- Select "Service" from the main menu.
Illustration 2 | g03477356 |
Available Calibrations in Cat® ET |
- Select "Calibrations" from the "Service" drop-down menu.
The following is a list of all possible calibrations from the "Calibrations" drop-down menu. The specific calibrations will differ across sales models.
- "PRV - Main Pump Flow Limitation Pressure Calibration"
- "PRV - Power Shift Pressure Calibration"
- "Modulation Input Calibrations"
- "PRV - Variable Relief Pressure Calibration"
- "PRV - Attachment Valves Pressure Calibration"
- "Engine Speed Calibrations"
- "PRV - Solenoid Pressure Calibrations"
- "PRV - Main Pump Flow Limitation Pressure Calibration"
- From the "Calibrations" drop-down menu, select the desired calibration procedure.
- Follow the instructions that are provided on the screen.
Note: Additional screens will be displayed in order to notify the operator of the status of the calibration procedure.
Typical Calibration Screens
Illustration 3 | g03477363 |
Typical Calibration Warning |
Illustration 4 | g03477377 |
Typical Select List for Solenoid Valve Calibration |
Illustration 5 | g03477396 |
First Step of PRV Calibration |
Hydraulic Cab Riser (HCR) Calibration
This calibration procedure is for the Hydraulic Cab Riser (HCR) angle sensor (if equipped), using the Monitoring System display located in the cab. This calibration is required if the machine ECM or HCR front angle sensor have been replaced.
- Ensure that HCR configuration is set as "Installed". Using the machine monitor display, go to MAIN MENU ⇒ SERVICE ⇒ CONFIGURATIONS ⇒ ATCH PARTS INST ⇒ HCR.
Note: A password will be needed when service mode to chosen.
Illustration 6 | g03474703 |
HCR Configuration |
- Secure the machine for HCR angle sensor calibration.
- Lower stabilizers to raise the front and rear axles off the ground.
Show/hide tableIllustration 7 g03474618
Place HCR into the middle position
- Position the worktool on the ground.
- Raise the HCR to the middle position. Check the menu status to determine the middle position.
- Select, MAIN MENU ⇒ SERVICE ⇒ CALIBRATIONS ⇒ ANGLE SENSOR ⇒ FRONT ANGLE and press "OK" after confirmation.
Note: For calibration error code information, refer to Table 1.
Illustration 8 | g03474723 |
HCR Front Angle Calibration - Start |
- Slowly lower the HCR to the bottom position and press "OK".
Note: The snubbing system is canceled during calibration. If HCR hits the stop at maximum speed, the system will be greatly shocked and damage could result.
Illustration 9 | g03474736 |
HCR Front Angle Calibration - Set to Bottom Position |
- Slowly raise the HCR to the top position and press "OK".
Illustration 10 | g03474737 |
HCR Front Angle Calibration - Set to Top Position |
- After calibration has succeeded, confirm that the snubbing system works correctly. To go back to the previous screen, press the bottom back key.
Illustration 11 | g03474740 |
HCR Front Angle Calibration - Succeeded |
- Adjust the snubbing system, if the shock is excessive, when the HCR reaches the stops. To adjust, navigate to MAIN MENU ⇒ SERVICE ⇒ CONFIGURATIONS ⇒ PARAMETER SETTING ⇒ E-FENCE on the machine monitor.
Illustration 12 | g03474742 |
HCR Up / Down Gain - To Adjust Snubbing |
"GAIN1" = angle gain. 5 is Narrow (Heavy), 0 is Normal and -5 is Wide (Light).
"GAIN2" = rate gain. 5 is High (Heavy), 0 is Normal and -5 is Low (Light).
.
Error Codes That Can Be Activated during HCR Angle Sensor Calibration     | ||
---|---|---|
Error Codes (1)     | Error Descriptions     | Recommended Response / Possible Causes     |
0002     | Active Diagnostic Code     | Resolve the problem that is causing the active diagnostic code. Then, retry the calibration.     |
0003     | Other Calibration or Test Is Running     | Only one diagnostic test can be active. Wait for diagnostic testing to complete. Or key OFF once and restart calibration.     |
0004     | Calibration Active By Another ECM     | Calibration procedures are trying to run on the monitor and Cat ET at the same time. Or a calibration was started, then canceled and another calibration was started too quickly. Restart calibration after other calibration is performed. Or key OFF once and restart calibration.     |
0005     | Loss Of Calibration Interlock     | Check and resolve any active Cat Data Link (CDL) diagnostic codes. Check the CDL and monitor connections. Cycle the key start switch from ON to OFF, wait 10 seconds, then turn the key start switch to ON and rerun the calibration procedure.     |
0006     | Calibration Aborted By Tool / Monitor     | Service technician interrupted or canceled the calibration. Wait at least 15 seconds before performing another calibration.     |
0007     | Service Tool Is Not Supported     | The machine monitor software does not support the calibration. Download and install the correct flash file and restart the calibration.     |
0009     | Calibration Aborted By ECM     | Normally another error code will also be active indicating the cause of the cancellation. Check that the (HCR) configuration is correct and restart the calibration.     |
000A     | Calibration Failure     | Failure of the calibration due to engine speed anomaly. After resolving the anomaly restart the calibration.     |
000B     | Calibration Not Supported     | Incorrect / early version of monitor software does not support the calibration. Verify that the correct flash file is downloaded into the monitor or machine ECM. After the ECM has been flashed, cycle the key start switch from ON to OFF. Wait 10 seconds, then turn the key start switch to ON and rerun the calibration.     |
0100     | Out-Of-Range Low     | HCR cylinder moves and/or the HCR angle sensor signal is below the allowed minimum range. Perform calibration again.     |
0101     | Out-Of-Range High     | HCR cylinder does not move and/or the angle sensor signal is outside the allowed maximum range. Perform calibration again.     |
0102     | Invalid Command     | Incorrect calibration procedure. Perform calibration again.     |
0103     | Calibration Value Out-of-Range     | The duty cycle from the HCR angle sensor is out-of-range. Ensure that the HCR is moved to the position indicated by the instructions in the calibration procedure. Perform calibration again.     |
1010     | Engine Stopped     | Troubleshoot the engine shutdown. Restart the engine and perform calibration again.     |
1014     | Low Engine Idle Speed     | Engine idle speed is below 1014 rpm. Correct the engine idle speed and perform calibration again.     |
1015     | Abnormal Engine Speed     | Engine speed sensor shows an abnormal state. Check, reinstall and/or replace the speed sensor. Perform calibration again.     |
1034     | Excessive Engine Speed     | Calibration not possible due to high engine rpm. Error occurs after the amount of time given to lower engine rpm is exhausted. Correct the engine speed issue and perform calibration again.     |
1035     | Engine Speed Response Delay     | Error indicates that engine rpm is not responding or is slow. Perform troubleshooting for an abnormality engine speed command signal. Perform calibration again.     |
1040     | Excessive Load During Calibration     | Error indicates that load was applied during calibration. Ensure that the hydraulic lock lever is in the locked position and perform calibration again.     |
1054     | Hydraulic Oil Temperature Too High     | Hydraulic oil temperature reading is higher than the maximum required for calibration. Typical maximum required hydraulic oil temperature is 90° C (194° F).     |
1074     | Engine Output Speed Is Less Than The Minimum Engine Speed     | Engine high idle speed signal is below the required threshold. Check that the electrical wiring from the engine speed sensor to the ECM is correct and perform calibration again.     |
( 1 ) | Error codes are hexadecimal numbers. |